The operator touches the container and immediately records a value.
Wait for the balance and container to equilibrate; document the handling condition and repeat only under the defined method.
Module 4 · Practice
Make a stable, traceable water or container weighing while separating the balance reading from the conclusion you draw.
Objective
A balance reading is affected by the instrument, the support, air movement, temperature, static, vibration, and how the sample or container is handled. Sartorius operating instructions emphasize allowing samples to equilibrate to the balance environment and minimizing convection around a container. A draft shield can help, but it cannot repair a wet bench, an unstable installation, or a container warmed by your hand. Begin by reading the exact manual for the model in front of you; display resolution is not the same as measurement accuracy.
Tare is a bookkeeping operation that sets the displayed container mass to zero. It does not erase the physical container mass or prove the zero is correct. For traceability, record the empty-container reading if your method permits, then the gross reading, and calculate net mass as gross minus tare. If the balance prints or stores a result, preserve that record with the balance ID and time.
Use a clean, dry container and a small amount of room-temperature water. Do not weigh unknown powders, volatile chemicals, or biological material in this lesson. Water can evaporate and the container can develop droplets, so the exercise teaches observation and records rather than a universal density or accuracy claim. Close the draft shield, wait for stability, and avoid leaning on the bench while the value settles.
Worked example
The empty container reads 18.432 g and the gross container with water reads 28.917 g. Net mass is 28.917 g − 18.432 g = 10.485 g.
Materials
Safety & stop conditions
Method
Confirm the balance ID, level or status indicator, clean pan, stable bench, and manual-specified warm-up or environmental conditions. Record anything unusual.
Use a dry container at room conditions and handle it by a clean outside surface. Let the display settle before recording the empty mass.
If using tare, place the empty container, close the shield, wait for stability, and press tare according to the manual. Where possible, retain the pre-tare value too.
Add water without touching the pan or splashing. Close the shield, wait for the stability indicator, and record the net reading, units, timestamp, and any drift.
Remove the container, inspect for droplets, and leave the pan clean and dry. If the zero does not return as expected, stop and investigate rather than forcing a correction.
Checkpoints
Common mistakes
Wait for the balance and container to equilibrate; document the handling condition and repeat only under the defined method.
Treat it as incomplete raw data and recover the metadata before using the value.
Close the shield, remove drafts and vibration, and investigate temperature or evaporation before deciding whether to repeat.
Use the model specification and verification evidence; resolution alone cannot establish uncertainty.
Practice
If a suitable balance is available under its approved instructions, weigh one dry container and then the same container with water; repeat after removing and replacing it. Without equipment, use a tare of 18.432 g and gross readings of 28.917 g and 28.919 g after replacing the same container. Calculate the two net readings and audit the calculation trail. These are invented practice values, and the two readings are not independent samples of water.
Self-checks
Answer: It subtracts or zeroes the container contribution in the balance display; it does not prove the instrument is calibrated.
Answer: To reduce air movement and convection that can cause an unstable or biased display.
Answer: The calculation trail and a way to audit container identity and subtraction are missing.
Answer: Only if the instrument specification and method support them; display resolution is not a substitute for uncertainty evaluation.