The tip is immersed deeply and liquid rises into the shaft.
Stop, follow the instrument cleaning guidance, and repeat with the manual’s immersion depth. Do not continue using a contaminated pipette.
Module 4 · Practice
Choose a pipette for the volume and liquid, then transfer water with a repeatable hand position and documented checks.
Objective
A pipette does not make every volume equally trustworthy. First write the target volume, its unit, and the liquid. Select a device whose published operating range contains that volume; near the middle of a range is often easier to control, but the manufacturer’s instructions and your method decide what is acceptable. Use a compatible, undamaged tip. A transfer from a clean water stock into a clean tube is a bounded practice exercise. It does not qualify you to handle samples, volatile liquids, viscous liquids, corrosives, or biological material.
Forward pipetting uses the first plunger stop to aspirate the set volume and the second stop to empty the tip. The technique depends on a slowly returning plunger, a nearly vertical tip, controlled immersion, and a consistent dispense against the vessel wall. Pre-wetting the tip can reduce evaporation effects in air-displacement systems; the Eppendorf operating manual recommends aspirating and dispensing the liquid one to three times before the measured uptake. The point is consistency, not a ritual number detached from the instrument manual.
Practice with water and a single planned volume. Put the source and destination at a comfortable height, label the destination before starting, and write down the pipette ID, set volume, tip type, operator, date, and any instrument-specific temperature or pre-wetting instruction. A useful checkpoint asks what you observed: bubbles, a droplet outside the tip, an unstable meniscus, or a missed stop. Each observation is data about the transfer, not an embarrassment to hide.
Worked example
A learner uses pipette P-02, set to 1.00 mL, with a compatible tip and clean water at room conditions. The first transfer shows no bubble; the second has a droplet on the outside of the tip before dispense.
Materials
Safety & stop conditions
Method
Write the target, unit, source identity, destination identity, and acceptance rule before touching the pipette. This turns a hand movement into a traceable measurement.
Confirm the range, fit a compatible tip, and set the volume without forcing the adjustment beyond its limits. Check that the tip is seated without cracking it.
With water, aspirate and dispense one to three times if the manufacturer’s instructions call for pre-wetting. Keep the hand motion slow enough to avoid bubbles.
Press to the first stop, immerse the tip to the manual’s recommended depth, and let the plunger return slowly while the pipette remains vertical. Pause briefly, then withdraw without scraping the wall.
Touch the tip to the receiving-wall, press to the first stop, wait for drainage, then press to the second stop and withdraw while still holding the plunger. Record observations and change the tip when the method requires it.
Checkpoints
Common mistakes
Stop, follow the instrument cleaning guidance, and repeat with the manual’s immersion depth. Do not continue using a contaminated pipette.
Treat that transfer as suspect, expel safely into the source if allowed, inspect the tip fit, slow the plunger, and record the event.
Reset and use the first stop for uptake; the second stop is the blow-out portion of forward dispensing.
Record the new ID and confirm that its range, tip, and method are appropriate before combining results.
Practice
If you have a permitted pipette whose published range includes 1.00 mL, set it to 1.00 mL and transfer water into three labeled tubes. Otherwise use the W1/W2 records in the worked example below and annotate the expected checkpoints without operating equipment. This is a technique exercise, not evidence that the pipette is calibrated.
Self-checks
Answer: The published range defines where the device is intended to operate; a target outside it has no justified volume claim.
Answer: It is used for the measured aspiration and the initial dispensing stroke; the second stop completes emptying the tip.
Answer: No. Mark that transfer suspect, follow the method’s repeat rule, and retain the original observation in the raw record.
Answer: It conditions the air cushion and can reduce evaporation-related volume effects, especially for small volumes; the instrument manual controls the exact practice.
Sources & further reading