A certificate is treated as proof of current fitness.
Use it as part of the evidence, then verify current performance under the intended conditions when the method requires it.
Module 5 · Concept
Separate adjustment, evidence against a reference, and the decision that an instrument is suitable for a particular measurement.
Objective
Calibration compares an instrument indication with a reference and documents the relationship and uncertainty. Adjustment changes the instrument so it indicates differently; an adjustment may be part of a calibration service, but it is not the same evidence. Verification asks whether the instrument or method meets a stated requirement at the time and conditions of use. Validation asks whether the complete method is fit for its intended purpose. A manufacturer’s calibration certificate can support a claim, but it does not automatically prove that today’s method is suitable for every volume, temperature, or sample type.
Start with the decision. If the instrument will distinguish 9.8 from 10.2 units, a check designed only for a broad range may be irrelevant. Define the measurand, range, reference, conditions, acceptance criterion, and response to failure before collecting a check result. The criterion must come from a method, manufacturer specification, contract, regulation, or documented scientific rationale. Never copy a tolerance from an unrelated instrument.
A useful record links instrument ID and software or configuration to reference identity, reference value and uncertainty, date, operator, conditions, raw indication, calculation, criterion, and decision. ISO/IEC 17025 frames competence and consistent laboratory operation; the practical lesson is to make the evidence auditable even when you are working on a small project. A failed verification does not prove every earlier result is wrong, but it does create a question about the affected interval and requires a documented investigation.
Worked example
A method needs a balance indication near 10 g and defines an allowable difference of 0.010 g. A check reference is 10.000 g with a documented uncertainty; the balance reads 10.006 g.
Materials
Safety & stop conditions
Method
Write the measurand, range, units, environment, and decision the measurement will support.
Locate the applicable manufacturer specification, method, certificate, regulation, or documented project criterion. Record its version or date.
Choose a reference and check point that exercise the relevant range and uncertainty. State why the check represents the intended use.
Record the raw indication, correction or difference, uncertainty information, criterion, and accept/investigate decision. Do not back-fit a criterion after seeing the result.
If the check fails, identify the affected measurements, place the instrument in the method’s defined status, and document review or repeat actions.
Checkpoints
Common mistakes
Use it as part of the evidence, then verify current performance under the intended conditions when the method requires it.
Describe whether the instrument was compared, changed, or judged against a criterion.
Find a justified criterion or state that the check cannot support a pass/fail conclusion.
Preserve the failed record, document the action, and assess affected data before repeating.
Practice
Use this supplied record: a 10.000 g reference gives 10.006 g and a 50.000 g reference gives 50.021 g. The predeclared comparison limits are ±0.010 g and ±0.050 g respectively. Decide whether the record supports the intended 10 g water-weighing exercise, and identify missing evidence.
Self-checks
Answer: No. Adjustment changes indication; calibration is the documented comparison with a reference and uncertainty.
Answer: Before the check, from a relevant method, specification, requirement, or documented scientific rationale.
Answer: No. The check point and uncertainty may not represent the smaller measurement.
Answer: The raw result, criterion, instrument status, affected-data assessment, action, and later verification record.
Sources & further reading